Home / Products / Light Meters 2.0 / X12 2.2.2
|
X12 2.2.2
|
| Description: |
|
Universal Optometer for Multi-Purpose Light Measurement The X12 Optometer device offers a RS232 interface and multi- application capability. With its compatibility to detectors with ITT connector, the X1 2 offers various kinds of measurement modes. One unique feature is the capability to operate heads with up to 4 sensors. Its simple operation and functionality makes the X1 1 to an ergonomically device. Optometer and associated light detector calibration is traceable to the ISO EN 17025 (Reg. No. DKD-K-10601-03) accredited Gigahertz-Optik calibration laboratory for optical radiation quantities.
Features:
- Compatible to all GO detectors with ITT connector
- CW, Dose and Color measurement modes
- RS232 interface
- Adjustable integration time
- Windows based OS-X1 1 Software available
- Inexpensive
For Light Measurement including:
- Luminous flux/ Radiant Power
- Luminous intensity / Radiant Intensity
- Luminance / Radiance
- Illuminance / Irradiance
- Laser Power / Laser Stray Light
- Photosynthetically Active Radiation (PAR)
- Color measurement
|
|
|
| Specification: |
|
|
Detector heads |
All available detectors with -4 type data connector. See chapter light detectors to select the detector head for your application |
|
Measurement modes |
CW light measurement/ Dose measurement/ Pulse synchronization/ Light source data logger/ Color |
|
Measurement range |
8 decade stepped gain ranges from 0,1pA- 200μA with Auto/ Manual range switch |
|
Slew rate |
3ms- 30ms |
|
Integration time |
1ms- 1s |
|
Sampling rate |
1ms |
|
Signal conversion |
0.166 Hz to >300MHz |
|
Display |
LCD- display 97x32 pixel with LED backlight illumination |
|
Interface |
RS232 |
|
Operating Temperature |
10 to 40°C (50 to 104°F) (75 % rel. H, non condensing) |
|
Storage Temperature |
0 to 50°C (32 to 122°F). |
|
Dimension |
145 x 63 x 30 mm / 5.7 x 2.5 x 1.2 in |
|
Weight |
150 g/ 0.33 lb |
|
Power supply |
2 x battery size AA (2.2 to 3.2V | |
|
|
Datasheet printversion |
© 2008,2009 Gigahertz-Optik GmbH - All rights reserved | IMPRINT | CONTACT US | SITEMAP