The X4-DE-UN contains the diode array and its electronics as well as the electronics for the integral detectors. Both units are remote controlled via the USB interface. The diode array spectrometer in a compact Czerny Turner design is equipped with a 2048 pixel CCD array. A SMA fiber optic input with collimating, focusing mirror and diffractional grating enable low light level and high resolution light measurement. A flexible light guide connects the diode array spectrometer to the BTS detector head. The integral detector electronics consists of a high sensitivity analog amplifier with switchable gain over a 0.1-100pA to 0.2-200µA measurement range. The 3 ms rise time allows peakvalue data logging for short term mode measurements of LEDs. A remote controlled switch at the signal input permits operation of up to four detectors in multiplexing mode. The X4-DE-UN in combination with the X4-BTS detector head is the right tool for general purpose light analysis. In combination with optional light source amplifiers, stepping motor controller and peltier temperature controller the X4-DE-UN forms the basis for a light source controller set-up for complete control and analysis of discharge and incandescent lamps as well as for LEDs, OLEDs and flat panel displays. Windows based software is supplied with the X4-DE light analyzer which enables stand alone use as well as system integration. System integration is supported by the DLL based structure. The X4-DE-UN is offered in either a bench-top or rack-mount housing. |
X4-DE-UN |
|
X4-DE-UN-RM |
|
| Specification: |
X4-DE-UN Detector Electronic Unit Specifications
|
Integral Detector Electronics: |
|
Detector Input |
-4 type connector |
|
Signal amplifier |
Current to Voltage |
|
Signal range |
200.0 pA to 200.0 µA with 7 gain ranges |
|
Max. resolution |
0.1 pA in 200.0 pA Range |
|
Gain linearity error |
0.2 % of reading;+ 0.05 % of range; 23° +/- 5°C; with offset compensation |
|
Rise time 10 - 90 % |
30 ms (gain 1 to 2), 3 ms (gain 3 to 6) |
|
AD converter |
12 bit |
|
Diode Array: |
|
Optical Input |
SMA connector |
|
Geometrical design |
Symmetrical Czerny-Turner, 75 mm focal length |
|
Wavelength Range |
300 to 1050nm ( UVNIR ) |
|
Optical Resolution |
2.1nm |
Pixel Resolution |
Approximate 0.5 nm |
|
Stray Light |
< 0.1% |
|
Detector |
CCD linear array, 2048 pixels |
|
Signal/Noise |
250 : 1 |
|
Integration time |
2 to 10000 millisecond |
|
AD converter |
14 bit, 2 MHz |
|
General |
|
Remote Interface |
USB version 1.1, 12 Mbit |
|
Operating Temperature |
10 to 40° C (50 to 104° F) (75 % rel. H, non-condensing). Storage Temperature: 0 to 50°C (32 to 122° F). |
|
Dimensions/Weight |
Bench-top version 170 x 110 x 71 mm. |
|
Power |
5VDCUSB power |
|
|
Additional Information  printversion |