• Calibration of the Absolute Spectral Irradiance Sensitivity [A/(W/m2) nm] of Irradiance Detectors by the Gigahertz-Optik ISO 17025 Accredited Calibration Laboratory for Optical Radiation Quantities
• Monochromatic source signal (single monochromator with deuterium or tungsten lamp and collimating optic)
• Beam diameter larger than the detection area of the probe)
• Comparison of test device signal (A) against ISO/IEC/EN 17025 traceable calibration standard detector
• Factory Type Calibration Certificate
Ordering Information:
• KDW-S2-01: Spectral Irradiance Calibration at one wavelength from 200 to 1100 nm
• KDW-S2-02: at one additional Wavelength to KDW-S2-01
• KDW-S2-03: Spectral Irradiance Calibration in 10 nm steps from 200 to 400 nm
• KDW-S2-04: Spectral Irradiance Calibration in 10 nm steps from 350 to 1100 nm
• KDW-S2-05: Spectral Irradiance Calibration in 10 nm steps from 200 to 1100 nm
• KDW-S2-RP: Set-up charge for non-standard product

Additional information in
Calibration_Service.pdf (0,4 MB)